Oxygen sensor for Hamilton 396008, 396009, Galileo, Raphael, Arabella, Aladdin, 51254, G5, Datascope Multinex Analyzer 0600-00-0002, Hudson 5556
Oxygen Fuel Cell for Hamilton HM01, Hudson 5569
Sensor Specifications & Compatibility | |
Catalog Number | E61912 |
Output Signal in Air | 170A - 230A no temperature compensation |
Response Time T 90% | < 12 seconds |
Nominal Sensor Lifetime | 250 000 % volume Oxygen hours |
Electrical Interface | Gold plated slip rings |
Compatibility: | Hamilton HM-01, Veolar, Amadeus, Arabella |
Novacare | |
Hudson 5568, C1, C-01 | |
Oxequip Oxitron 3250 | |
Vickers incubator | |
Hamilton HM-11 | |
SS01 (SS-01) | |
MAX19 | |
PSR-11-33-1 | |
10-103-00 | |
PT01 | |
Teledyne T1 A-37016 | |
Medical Oxygen Sensors (Fuel Cells) general specifications: | |
Measurement Range: | 0-100 % oxygen |
Accuracy and Repeatability: | < 1 % vol. O2when calibrated at 100 % O2 |
Zero Ofset: | < 0.5 % vol. O2in 100% N2, applied 5 minutes |
Linearity Error: | < 3 % relative |
Cross Interference: | < 0.5 % vol. O2response to: |
10 % CO2balance N2 | |
80% N2O balance N2 | |
7.5% Halothane balance N2 | |
7.5 % Isofurane balance N2 | |
7.5 % Enfurane balance N2 | |
9% Sevofurane balance N2 | |
20% Desfurane balance N2 | |
Infuence of Humidity: | - 0.03 % rel. per % RH at 25C |
Infuence of Pressure: | proportional to change in oxygen partial |
pressure | |
Infuence of Mechanical Shock: | < 1% relative after a fall from 1m |
Operating Temperature: | 0 to 50C |
Temperature Compensation: | built-in NTC compensation (see below) |
Efect of Temperature | between +5 C and +25 C: 3 % relative error |
Compensation (steady state): | between 0 C and +50 C: 8 % relative error |
Operating Humidity: | 0-99 % RH non-condensing |
Long Term Output Drift: | < 1 % vol oxygen per month |
typically < - 15 % relative over lifetime | |
Storage Temperature: | -20 to +50 C |
Recommended Storage: | +5 to +15 C |
Recommended Load: | 10 kOhms |
Warm-Up Time: | < 30 minutes, after replacement of sensor |
Weight: | approximately 28 grams |
approximately 43 grams E61912 series | |
Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003 | |
specifications are applicable at standard conditions: 1013 hPa, 25C dry ambient air |