Oxygen sensor for Puritan Bennet Ventilators, PB840, PB700, PTS-2000
Click To Enlarge

Oxygen sensor for Ventronics 5528 and 5529, Hudson 5565, 5550 and 5568

Oxygen Fuel Cell for Ventronics, Hudson 5568

Sensor Specifications & Compatibility    
Catalog Number E61550  
Output Signal in Air 170A - 230A no temperature compensation  
Response Time T 90% < 12 seconds  
Nominal Sensor Lifetime 250 000 % volume Oxygen hours  
Electrical Interface Flying leads with pin-connectors  
Compatibility: Hudson 5565 (Analyzer 5525 and 5552)  
  Hudson 5550 and 5568  
  Ventronics 5528 and 5529  
  MAX18  
  10-103-01  
  PT02 (PT-02)  
  Teledyne C1R, B-39022  
  PSR-11-33-1  
 
Medical Oxygen Sensors (Fuel Cells) general specifications:  
Measurement Range: 0-100 % oxygen  
Accuracy and Repeatability: < 1 % vol. O2when calibrated at 100 % O2  
Zero Ofset: < 0.5 % vol. O2in 100% N2, applied 5 minutes  
Linearity Error: < 3 % relative  
Cross Interference: < 0.5 % vol. O2response to:  
  10 % CO2balance N2  
  80% N2O balance N2  
  7.5% Halothane balance N2  
  7.5 % Isofurane balance N2  
  7.5 % Enfurane balance N2  
  9% Sevofurane balance N2  
  20% Desfurane balance N2  
Infuence of Humidity: - 0.03 % rel. per % RH at 25C  
Infuence of Pressure: proportional to change in oxygen partial  
  pressure  
Infuence of Mechanical Shock: < 1% relative after a fall from 1m  
Operating Temperature: 0 to 50C  
Temperature Compensation: built-in NTC compensation (see below)  
Efect of Temperature between +5 C and +25 C: 3 % relative error  
Compensation (steady state): between 0 C and +50 C: 8 % relative error  
Operating Humidity: 0-99 % RH non-condensing  
Long Term Output Drift: < 1 % vol oxygen per month  
  typically < - 15 % relative over lifetime  
Storage Temperature: -20 to +50 C  
Recommended Storage: +5 to +15 C  
Recommended Load: 10 kOhms  
Warm-Up Time: < 30 minutes, after replacement of sensor  
Weight: approximately 28 grams  
  approximately 43 grams E61912 series  
Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003  
specifications are applicable at standard conditions: 1013 hPa, 25C dry ambient air

 

 

O2 Sensor for Ventronics 5528 and 5529, Hudson 5565, 5550, 5568



  • Item #: B226-218
  • Manufacturer: WestPrime Healthcare
  • Manufacturer's Product Number: OOM107-2 / E61550
  • Condition: New

O2 Sensor for Ventronics 5528 and 5529, Hudson 5565, 5550, 5568

Price: $107.50
* Marked fields are required.
Availability: Back Ordered
Qty: *
Reviews (0) Write a Review
No Reviews. Write a Review