Sensor Specifications & Compatibility | ||
Catalog Number | E61800 | |
Output Signal in Air | 10mV 15.5mV | |
Response Time T 90% | < 15 seconds | |
Nominal Sensor Lifetime | 900 000 % volume Oxygen hours | |
Electrical Interface | Monojack | |
Compatibility: | Maxtec MaxO2+AE, OM-25AE, OM-25ME | |
MaxBlend, MaxVenturi MAX250E (MAX-250E) | ||
CareFusion (Cardinal, Viasys) Sentry Blender | ||
Newport Medical E360 & E500 ventilators | ||
Teledyne R29MED , C-70592 | ||
Viasys Sentry Blender, 66045 | ||
10-103-08 | ||
Fluke VT Mobile | ||
PSR-11-917-J1 | ||
Notes: External Type sensor, see E61810 for Internal sensor. | ||
Medical Oxygen Sensors (Fuel Cells) general specifications: | ||
Measurement Range: | 0-100 % oxygen | |
Accuracy and Repeatability: | < 1 % vol. O2when calibrated at 100 % O2 | |
Zero Ofset: | < 0.5 % vol. O2in 100% N2, applied 5 minutes | |
Linearity Error: | < 3 % relative | |
Cross Interference: | < 0.5 % vol. O2response to: | |
10 % CO2balance N2 | ||
80% N2O balance N2 | ||
7.5% Halothane balance N2 | ||
7.5 % Isofurane balance N2 | ||
7.5 % Enfurane balance N2 | ||
9% Sevofurane balance N2 | ||
20% Desfurane balance N2 | ||
Infuence of Humidity: | - 0.03 % rel. per % RH at 25ðC | |
Infuence of Pressure: | proportional to change in oxygen partial | |
pressure | ||
Infuence of Mechanical Shock: | < 1% relative after a fall from 1m | |
Operating Temperature: | 0 to 50ðC | |
Temperature Compensation: | built-in NTC compensation (see below) | |
Efect of Temperature | between +5 ðC and +25 ðC: 3 % relative error | |
Compensation (steady state): | between 0 ðC and +50 ðC: 8 % relative error | |
Operating Humidity: | 0-99 % RH non-condensing | |
Long Term Output Drift: | < 1 % vol oxygen per month | |
typically < - 15 % relative over lifetime | ||
Storage Temperature: | -20 to +50 ðC | |
Recommended Storage: | +5 to +15 ðC | |
Recommended Load: | 10 kOhms | |
Warm-Up Time: | < 30 minutes, after replacement of sensor | |
Weight: | approximately 28 grams | |
approximately 43 grams E61912 series | ||
Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003 | ||
specifications are applicable at standard conditions: 1013 hPa, 25ðC dry ambient air |