Envitec oxygen sensors by Honeywell are the best in the medical industry for reliability, high precision, fast response time and excellent signal stability. 13mV-16mV output at 21% air. 3 pin Molex connector Oxygen Sensor Cell replacement for the following equipment: Puritan Bennett 840 Ventilator Puritan Bennett 700 Series Ventilator Maquet Servo 300 Ventilator Maquet Servo 900C Ventilator Carefusion Avea Ventilator Hamilton Aladdin Infant Flow System Hamilton Arabella Infant Flow System Hamilton Galileo Ventilator Hamilton Raphael Ventilator Carefusion Infant Flow System.

Made in Germany.

Sensor Specifications & Compatibility  
Catalog Number E61840  
Output Signal in Air 13mV - 16mV temperature compensated  
Response Time T 90% < 12 seconds  
Nominal Sensor Lifetime 1 000 000 % volume Oxygen hours  
Electrical Interface 3 pin molex  
Compatibility: Nellcor Puritan Bennett Malinckrodt  
  Mod. 740, 760, 840  
  G-0603400  
  G-062010-00  
  4-072214-00  
  G061143-00  
  G062009-00  
 
Medical Oxygen Sensors (Fuel Cells) general specifications:  
Measurement Range: 0-100 % oxygen  
Accuracy and Repeatability: < 1 % vol. O2when calibrated at 100 % O2  
Zero Ofset: < 0.5 % vol. O2in 100% N2, applied 5 minutes  
Linearity Error: < 3 % relative  
Cross Interference: < 0.5 % vol. O2response to:  
  10 % CO2balance N2  
  80% N2O balance N2  
  7.5% Halothane balance N2  
  7.5 % Isofurane balance N2  
  7.5 % Enfurane balance N2  
  9% Sevofurane balance N2  
  20% Desfurane balance N2  
Infuence of Humidity: - 0.03 % rel. per % RH at 25ðC  
Infuence of Pressure: proportional to change in oxygen partial  
  pressure  
Infuence of Mechanical Shock: < 1% relative after a fall from 1m  
Operating Temperature: 0 to 50ðC  
Temperature Compensation: built-in NTC compensation (see below)  
Efect of Temperature between +5 ðC and +25 ðC: 3 % relative error  
Compensation (steady state): between 0 ðC and +50 ðC: 8 % relative error  
Operating Humidity: 0-99 % RH non-condensing  
Long Term Output Drift: < 1 % vol oxygen per month  
  typically < - 15 % relative over lifetime  
Storage Temperature: -20 to +50 ðC  
Recommended Storage: +5 to +15 ðC  
Recommended Load: 10 kOhms  
Warm-Up Time: < 30 minutes, after replacement of sensor  
Weight: approximately 28 grams  
  approximately 43 grams E61912 series  
Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003  
specifications are applicable at standard conditions: 1013 hPa, 25ðC dry ambient air

 

O2 sensor for Puritan Bennet Ventilators, PB840, PB700, PTS-2000

4-072214-00 / OXYGEN SENSOR PURITAN BENNETT 840 

OOM202

  • Item #: B226-216
  • Manufacturer: WestPrime Healthcare
  • Manufacturer's Product Number: OOM202 / E61840
  • Condition: New

O2 sensor for Puritan Bennet Ventilators, PB840, PB700, PTS-2000

Price: $155.00
* Marked fields are required.
Availability: Back Ordered
Qty: *
Reviews (0) Write a Review
No Reviews. Write a Review